PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-50050
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CUSTOMER REPORTS TO HAVE RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 326 MG/DL, WHICH WAS TREATED WITH INSULIN PUMP. DURING TROUBLESHOOTING, IT WAS FOUND THAT CUSTOMER WAS NOT EXPOSED TO MAGNETIC FIELD OR MRI. CUSTOMER DOES NOT USE SENSOR FEATURES. CUSTOMER WAS ABLE TO COMPLETE REWIND SEQUENCE. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723857 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |