5MM MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00893
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 20, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. THE FOLLOWING REPAIR AND SERVICE DIAGNOSTIC CODES WERE IDENTIFIED: COATING PEELING OFF HANDLE, REPLACED HANDLE. THE FOLLOWING WAS OBSERVED: COATING PEELING OFF THE HANDLE. THE HANDLE WAS REPLACED. THIS DOES CONFIRM THE ALLEGED FAILURE MODE OF INSULATION DAMAGE. PROBABLE ROOT CAUSE: POOR AUTOCLAVE RELIABILITY, INCORRECT STERILIZATION/REPROCESSING PROCEDURE, HANDLING PROCEDURES, CONTACT FORCES, PRODUCT USED BEYOND DEFINED USEFUL LIFE. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723331 | 5MM MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 0541370D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |