FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR HANDLE 33CM

MDR report key: 4241785 · Received November 10, 2014

Report

Report Number
0002936485-2014-00893
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 20, 2014
Report Date
October 20, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. THE FOLLOWING REPAIR AND SERVICE DIAGNOSTIC CODES WERE IDENTIFIED: COATING PEELING OFF HANDLE, REPLACED HANDLE. THE FOLLOWING WAS OBSERVED: COATING PEELING OFF THE HANDLE. THE HANDLE WAS REPLACED. THIS DOES CONFIRM THE ALLEGED FAILURE MODE OF INSULATION DAMAGE. PROBABLE ROOT CAUSE: POOR AUTOCLAVE RELIABILITY, INCORRECT STERILIZATION/REPROCESSING PROCEDURE, HANDLING PROCEDURES, CONTACT FORCES, PRODUCT USED BEYOND DEFINED USEFUL LIFE. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723331 5MM MONOPOLAR HANDLE 33CM ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 0541370D

Patients

Seq Age Sex Outcome Treatment
1