5MM MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00890
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 20, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM MONOPOLAR HANDLE 33CM, DAMAGE TO THE INSULATION WAS CONFIRMED. VISIBLE DINGS AND SCRATCHES WERE SEEN THROUGHOUT THE SHAFT. ALSO, IT WAS NOTICED THAT THIS UNIT COULD HAVE POSSIBLY BEEN REPAIRED BY A THIRD PARTY IN THE PAST. THE 5MM MONOPOLAR HANDLE 33CM IS MISSING ETCHING AT THE PROXIMAL END OF THE SHAFT, WHICH CONSIST OF LOT NUMBER, CE MARK, US PAT NO AND REFERENCE NUMBER. IN ADDITION, A PIECE OF THE DISTAL TIP OF THE SHAFT WAS BROKEN OFF AND WAS NOT RECEIVED WITH THE HANDLE. THE 5MM MONOPOLAR HANDLE 33CM PASSED THE INSULATION INTEGRITY TEST. HOWEVER, IFU 10000401070 STATES "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE-THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE/S COULD BE USER MISHANDLING, IMPROPER STERILIZATION. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723312 | 5MM MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 0845186D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |