FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR HANDLE 33CM

MDR report key: 4241778 · Received November 10, 2014

Report

Report Number
0002936485-2014-00890
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 20, 2014
Report Date
October 20, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM MONOPOLAR HANDLE 33CM, DAMAGE TO THE INSULATION WAS CONFIRMED. VISIBLE DINGS AND SCRATCHES WERE SEEN THROUGHOUT THE SHAFT. ALSO, IT WAS NOTICED THAT THIS UNIT COULD HAVE POSSIBLY BEEN REPAIRED BY A THIRD PARTY IN THE PAST. THE 5MM MONOPOLAR HANDLE 33CM IS MISSING ETCHING AT THE PROXIMAL END OF THE SHAFT, WHICH CONSIST OF LOT NUMBER, CE MARK, US PAT NO AND REFERENCE NUMBER. IN ADDITION, A PIECE OF THE DISTAL TIP OF THE SHAFT WAS BROKEN OFF AND WAS NOT RECEIVED WITH THE HANDLE. THE 5MM MONOPOLAR HANDLE 33CM PASSED THE INSULATION INTEGRITY TEST. HOWEVER, IFU 10000401070 STATES "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE-THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE/S COULD BE USER MISHANDLING, IMPROPER STERILIZATION. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723312 5MM MONOPOLAR HANDLE 33CM ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 0845186D

Patients

Seq Age Sex Outcome Treatment
1