PENUMBRA COIL 400
Report
- Report Number
- 3005168196-2014-00802
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 3, 2014
- Report Date
- October 15, 2014
- Manufacturer
- PENUMBRA, INC.
- Product Code
- HCG
- PMA / PMN Number
- K120330
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSION: THIS DEVICE IS AVAILABLE FOR RETURN. A FOLLOW UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.
RESULT: THE PC400 COIL PUSHER ASSEMBLY WAS KINKED IN MULTIPLE LOCATIONS ALONG THE HYPO-TUBE. THE PUSHER ASSEMBLY WAS FRACTURED APPROXIMATELY 50.0 CM AND 56.0 CM FROM THE PROXIMAL END. THE PET LOCK WAS INTACT. THE COIL WAS DETACHED FROM THE PUSHER ASSEMBLY WITH THE PROXIMAL CONSTRAINT SPHERE INTACT. THE OUTER DIAMETER OF THE COIL, PROXIMAL CONSTRAINT SPHERE, PULL WIRE, AND DISTAL DETACHMENT TIP (DDT) WERE MEASURED WITHIN SPECIFICATION. THE INNER DIAMETER OF THE DDT WAS MEASURED WITHIN SPECIFICATION. CONCLUSION: THE COMPLAINT HAS BEEN EVALUATED. THE COMPLIANT INDICATES THAT A KINK WAS FOUND IN THE PC400 COIL DELIVERY WIRE DURING THE INSERTION THROUGH A PXSLIM. THE COMPLAINT ALSO STATES THAT THE DELIVERY PUSHER WAS PULLED SLIGHTLY AND THE COIL WAS CONFIRMED DETACHED IN THE PXSLIM. EVALUATION OF THE RETURNED PRODUCT REVEALED MULTIPLE KINKS IN THE PUSHER ASSEMBLY. THESE KINKS APPEAR TO BE FROM THE PACKAGING OF THE PRODUCT TO BE RETURNED FOR EVALUATION AND WERE NOT MENTIONED IN THE COMPLAINT REPORT. THE PUSHER ASSEMBLY WAS FRACTURED IN TWO LOCATIONS IN THE PROXIMAL END OF THE HYPO-TUBE. THESE FRACTURES COULD BE ASSOCIATED WITH THE COIL DETACHING INSIDE THE PXSLIM; HOWEVER, THE ROOT CAUSE OF THIS COMPLAINT CANNOT BE DETERMINED. THESE DEVICES ARE 100% FUNCTIONAL TEST DURING PROCESS INSPECTION. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS.
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE BASILAR ARTERY (BA) USING A PENUMBRA COIL 400. DURING THE PROCEDURE, WHILE ADVANCING THE THIRD PENUMBRA COIL 400, THE PHYSICIAN NOTICED THAT THE PUSHER ASSEMBLY OF THE COIL WAS KINKED. AS THE PHYSICIAN PULLED THE COIL, THE COIL UNINTENTIONALLY DETACHED INSIDE A PX SLIM DELIVERY MICROCATHETER. THE PX SLIM DELIVERY MICROCATHETER WAS REMOVED FROM THE PATIENT ALONG WITH THE DETACHED COIL. THE PX SLIM DELIVERY MICROCATHETER WAS FLUSHED WITH A SYRINGE TO REMOVE THE DETACHED COIL. THE PHYSICIAN CONTINUED THE PROCEDURE WITH THE SAME PX SLIM DELIVERY MICROCATHETER WITH A NEW PENUMBRA COIL 400. THERE WAS NO REPORT OF AN ADVERSE EFFECT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723311 | PENUMBRA COIL 400 | HCG, KRD | HCG | PENUMBRA, INC. | F38860 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |