FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4241731 · Received November 10, 2014

Report

Report Number
3004753838-2014-29525
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
June 1, 2014
Report Date
June 1, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE ERROR CODE. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON 10/14/2014. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(4) 2014 TO REPORT THAT ON (B)(6) 2014, RECEIVER VIBRATES CONTINUOUSLY. PATIENT STATED "RECEIVER DROPPED OUT OF THE CASE INTO TUB". AT THE TIME OF CONTACT THE PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723264 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649

Patients

Seq Age Sex Outcome Treatment
1 42 YR