FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4241731
·
Received November 10, 2014
Report
- Report Number
- 3004753838-2014-29525
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- June 1, 2014
- Report Date
- June 1, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE ERROR CODE. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON 10/14/2014. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE.
Description of Event or Problem · 1
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(4) 2014 TO REPORT THAT ON (B)(6) 2014, RECEIVER VIBRATES CONTINUOUSLY. PATIENT STATED "RECEIVER DROPPED OUT OF THE CASE INTO TUB". AT THE TIME OF CONTACT THE PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723264 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |