FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 4241581 · Received November 10, 2014

Report

Report Number
2032227-2014-50088
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE RESERVOIR LEAKED AROUND THE RUBBER GUM, THE BOTTOM GUM OF IT. CUSTOMER REPORTED HE ATTEMPTED TO BOLUS AND HE NOTICE THAT RESERVOIR LEAKED. CUSTOMER ATTEMPTED TO USE TWO RESERVOIRS FROM THE SAME LOT NUMBERS. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. CUSTOMER REPORTED THE LEAKS ONLY OCCURRED WHEN CUSTOMER ATTEMPTED TO BOLUS. CUSTOMER WAS ADVISED TO CLEAN RESERVOIR COMPARTMENT ON THE INSULIN PUMP. CUSTOMER WAS ADVISED TO PERFORM SELF TEST, TEST WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721645 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H9019563

Patients

Seq Age Sex Outcome Treatment
1 26 YR