FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4241561 · Received November 10, 2014

Report

Report Number
2032227-2014-49900
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 11, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED DISPLAY ANOMALY OR BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP HAD INTERMITTENT UP ARROW BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A BROKEN RESERVOIR TUBE LIP, SCRATCHED RESERVOIR TUBE WINDOW, A CRACKED RESERVOIR TUBE AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. IT WAS ALSO REPORTED THAT THE INSULIN PUMP HAS AN UNRESPONSIVE KEYPAD. CUSTOMER'S BLOOD GLUCOSE READING WAS 250 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722051 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 28 YR