FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4241556 · Received November 10, 2014

Report

Report Number
2032227-2014-49898
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 10, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO SKIPPING SCREEN, NO MOISTURE DAMAGE UNDER LCD SCREEN, ELECTRONIC ASSEMBLY OR MOTOR NOTED. THE INSULIN PUMP HAS CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN PUMP HAS BEEN EXPOSED TO MOISTURE DAMAGE. CUSTOMER REPORTED THAT MOISTURE CAN BE SEEN IN THE DISPLAY. THE BLOOD GLUCOSE LEVELS WERE NOT INCLUDED IN THE REPORT. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721993 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR