FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4241496 · Received November 10, 2014

Report

Report Number
2032227-2014-50660
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 15, 2014
Report Date
October 15, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION. FAILED BATTERY TEST ALARM DUE TO CORRODED BATTERY CAP CONTACT. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

NO BLOOD GLUCOSE LEVELS REPORTED. THE CUSTOMER REPORTED A FAILED BATTERY TEST ON THE INSULIN PUMP. NO ADDITIONAL INFORMATION WAS PROVIDED.