FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4241466 · Received November 10, 2014

Report

Report Number
2032227-2014-50517
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 12, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED UNABLE TO PRIME UNIT DURING PRIME/A33 TEST DUE TO FAULTY FORCE SENSOR RESISTOR. TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. THE INSULIN PUMP HAS MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP AND BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THE INSULIN PUMP DURING PRIMING. THE CUSTOMER'S BLOOD GLUCOSE WAS 278 MG/DL AND SHE TREATED WITH A BOLUS. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE INSULIN PUMP WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD OR MAGNETIC RESONANCE IMAGING MACHINE. THE CUSTOMER WAS NOT USING THE SENSOR FEATURE ON THE INSULIN PUMP. SHE WAS ABLE TO REWIND THE INSULIN PUMP AND WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722311 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523KNB

Patients

Seq Age Sex Outcome Treatment
1 10 YR