PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-50517
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 12, 2014
- Report Date
- October 12, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED UNABLE TO PRIME UNIT DURING PRIME/A33 TEST DUE TO FAULTY FORCE SENSOR RESISTOR. TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. THE INSULIN PUMP HAS MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP AND BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THE INSULIN PUMP DURING PRIMING. THE CUSTOMER'S BLOOD GLUCOSE WAS 278 MG/DL AND SHE TREATED WITH A BOLUS. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE INSULIN PUMP WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD OR MAGNETIC RESONANCE IMAGING MACHINE. THE CUSTOMER WAS NOT USING THE SENSOR FEATURE ON THE INSULIN PUMP. SHE WAS ABLE TO REWIND THE INSULIN PUMP AND WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722311 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523KNB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |