FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4241374 · Received November 10, 2014

Report

Report Number
1416980-2014-39725
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 14, 2014
Report Date
October 14, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND POWER ON SELF-TEST WERE PERFORMED. DURING VISUAL INSPECTION THE BATTERY WAS FOUND TO BE DAMAGED. THE CAUSE OF THE CONDITION WAS UNKNOWN. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A FIELD SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED BATTERY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723739 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1