PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-50079
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
CUSTOMER REPORTED THAT HIS INSULIN PUMP ALARMED OFF NO POWER. HE HAD JUST SHOWERED IN THE MORNING AND WAS TRYING TO DELIVER A BOLUS FOR BREAKFAST WHEN THE DEVICE SAID OFF NO POWER. INSERTING A NEW BATTERY DOES NOT RESOLVE THE ISSUE. CUSTOMER'S BLOOD GLUCOSE WAS 125 MG/DL. CUSTOMER STATED THE BATTERY CONTACTS WERE PRETTY DIRTY. CUSTOMER CLEANED THE CONTACTS AND INSERTER A NEW BATTERY AND THE DEVICE TURNED BACK ON. CUSTOMER WAS ADVISED TO MONITOR THE DEVICE. A REPLACEMENT BATTERY CAP WAS SENT. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723377 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |