FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4241209 · Received November 10, 2014

Report

Report Number
2032227-2014-50059
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD CONSTANT FAILED BATTERY TEST AFTER BATTERY INSERTION DUE TO FAULTY BATTERY TUBE. UNABLE TO TEST THE OPERATING CURRENTS, LOW BATTERY ALARM, OFF NO POWER ALARM AND SENSOR CALIBRATION ERROR ALARM DUE TO FAILED BATTERY TEST ALARM. UNIT HAD MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED FAILED BATTERY TEST. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. TROUBLESHOOTING WAS DONE. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. CUSTOMER MENTIONED ALSO ABOUT HIS SENSORS NOT STICKING AND GOT CALIBRATION ERRORS. CUSTOMER DECLINED TO TROUBLESHOOT. CUSTOMER STATED SHE BOUGHT NEW BATTERIES AND THE INSULIN PUMP WORKED AGAIN ADVISED TO DISCONTINUE THE USE OF THE INSULIN PUMP AND WAIT FOR THE REPLACEMENT TO AVOID ADDITIONAL ISSUES WITH THE SAME INSULIN PUMP. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723350 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAP

Patients

Seq Age Sex Outcome Treatment
1 35 YR