FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP

MDR report key: 4241206 · Received November 10, 2014

Report

Report Number
2032227-2014-50057
Event Type
Injury
Date Received
November 10, 2014
Date of Event
October 12, 2014
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. CUSTOMER'S BLOOD GLUCOSE WAS 460 MG/DL. TROUBLESHOOTING WAS DONE. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723349 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAS

Patients

Seq Age Sex Outcome Treatment
1 41 YR