FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4241137 · Received November 10, 2014

Report

Report Number
2032227-2014-49755
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 11, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY. MOISTURE DAMAGE WAS FOUND ON MOTOR ASSEMBLY. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THE INSULIN PUMP IS BLANK. CUSTOMER ALSO STATES THAT THE INSULIN PUMP IS EXPOSED TO MOISTURE UNDER THE LCD DISPLAY. THE BLOOD GLUCOSE READING IS 153 MG/DL .NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724225 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 30 YR