530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-49752
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 10, 2014
- Report Date
- October 12, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS AND MINOR SCRATCHED DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THEIR BLOOD GLUCOSE READING WAS HIGH. THE BLOOD GLUCOSE READING IS 153 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723682 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |