FDA Adverse Event Malfunction Summary report: N

MUSTANG?

MDR report key: 4240993 · Received November 10, 2014

Report

Report Number
2134265-2014-06762
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 16, 2014
Report Date
October 16, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
FGE
PMA / PMN Number
K110122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ARTERIO-VENOUS GRAFT OF THE FOREARM. A 6.0 X 40, 75CM MUSTANG¿ BALLOON CATHETER WAS USED TO DILATE THE LESION HOWEVER, THE BALLOON RUPTURED AT 18ATMS DURING THE FIRST INFLATION. THE BALLOON WAS COMPLETELY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722047 MUSTANG? CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - GALWAY H74939171060470 0016971551

Patients

Seq Age Sex Outcome Treatment
1 INFLATION DEVICE: EVEREST| INTRODUCER SHEATH: MEDIKIT 5FR| GUIDEWIRE: RADIFOCUS035