MUSTANG?
Report
- Report Number
- 2134265-2014-06762
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 16, 2014
- Report Date
- October 16, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- FGE
- PMA / PMN Number
- K110122
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ARTERIO-VENOUS GRAFT OF THE FOREARM. A 6.0 X 40, 75CM MUSTANG¿ BALLOON CATHETER WAS USED TO DILATE THE LESION HOWEVER, THE BALLOON RUPTURED AT 18ATMS DURING THE FIRST INFLATION. THE BALLOON WAS COMPLETELY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722047 | MUSTANG? | CATHETER, BILIARY, DIAGNOSTIC | FGE | BOSTON SCIENTIFIC - GALWAY | H74939171060470 | 0016971551 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INFLATION DEVICE: EVEREST| INTRODUCER SHEATH: MEDIKIT 5FR| GUIDEWIRE: RADIFOCUS035 |