DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2014-00351
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 31, 2014
- Report Date
- October 31, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- DIP
- PMA / PMN Number
- K911056
- Removal / Correction Number
- 2517506-09-24-2014-005C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED CUSTOMER COMPLAINTS OF IMPRECISION AND INACCURACY WITH PTN FLEX LOT FA5154. INTERNAL TESTING HAS CONFIRMED A LOW ABSORBANCE RANGE BETWEEN THE LEVELS 1 AND 5 CALIBRATORS WITH THIS LOT RESULTING IN THE POTENTIAL FOR A POSITIVE OR NEGATIVE BIAS AND IMPRECISION ACROSS THE PTN ASSAY RANGE. INTERNAL STUDIES DEMONSTRATED THE BIAS IS VARIABLE ACROSS THE ASSAY RANGE FROM 9-38%. THE HIGHEST DEGREE OF BIAS IS OBSERVED BELOW THE THERAPEUTIC RANGE OF 10-20 UG/ML (39.6-79.2 UMOL/L). THE REAGENT BLANK ABSORBANCE FOR PTN LOT FA5154 MAY ALSO GENERATE ABNORMAL REACTION FLAGS. SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION 14-63 DATED SEPTEMBER 2014 TO CUSTOMERS WHO HAD ORDERED PTN LOT FA5154 ADVISING THEM TO DISCONTINUE THE USE OF THE LOT. SIEMENS OFFERED A NO CHARGE REPLACEMENT WITH A NON-IMPACTED LOT OF PHENYTOIN FLEX REAGENT CARTRIDGES.
DISCREPANT PHENYTOIN (PTN) RESULTS WERE OBTAINED ON QC SAMPLES DURING A FAILED CALIBRATION WITH LOT FA5154. PATIENT RESULTS WERE NOT REPORTED TO PHYSICIANS. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS DELAYED, ALTERED OR PRESCRIBED ON THE BASIS OF THE FAILED PTN CALIBRATION. THERE IS NO INDICATION OF ADVERSE IMPACT TO PATIENTS DUE TO THE FAILED PTN CALIBRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 721888 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | PHENYTOIN FLEX® REAGENT CARTRIDGE | DIP | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |