FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4240809 · Received November 10, 2014

Report

Report Number
3004753838-2014-31662
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 1, 2014
Report Date
October 13, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

FOREIGN DISTRIBUTOR CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 ON PATIENT'S BEHALF TO REPORT INTERMITTENT OUT OF RANGE ON (B)(6) 2014. AT THE TIME OF CONTACT THE FOREIGN DISTRIBUTOR DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722120 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9438-01

Patients

Seq Age Sex Outcome Treatment
1