FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4240798
·
Received November 10, 2014
Report
- Report Number
- 3004753838-2014-31657
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION; HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. IT WAS DOWNLOADED AND REVIEWED ON 10/13/2014 CONFIRMING THE REPORTED EVENT OF INTERMITTENT OUT OF RANGE. A CAPA HAS BEEN INITIATED FOR THIS ISSUE.
Description of Event or Problem · 1
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT INTERMITTENT OUT OF RANGE SIGNAL ON (B)(6) 2014. AT THE TIME OF CONTACT THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722620 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR |