FDA Adverse Event Injury Summary report: N

FORTIFY ST DR

MDR report key: 4240753 · Received November 10, 2014

Report

Report Number
2938836-2014-17732
Event Type
Injury
Date Received
November 10, 2014
Date of Event
October 24, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

THE REPORTED FIELD EVENT OF PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON THE AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PREMATURE ERI. PATIENT CONDITION WAS NORMAL AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723179 FORTIFY ST DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD2235-40 NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention