FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240699 · Received November 10, 2014

Report

Report Number
2032227-2014-49679
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 10, 2014
Report Date
October 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A NO DELIVERY ALARM. CUSTOMER ALSO REPORTED HAVING RECENT HIGH BLOOD GLUCOSE LEVELS. BLOOD GLUCOSE LEVEL AT THE BEGINNING OF THE CALL WAS 207 MG/DL, BUT CUSTOMER REPORTED THAT BLOOD GLUCOSE LEVEL WAS 44 MG/DL LATER IN THE CALL. CUSTOMER REPORTED TREATING THE LOW WITH JUICE. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724146 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAH

Patients

Seq Age Sex Outcome Treatment
1 76 YR