FDA Adverse Event Malfunction Summary report: N

BATT F/BPL 14.4VDC

MDR report key: 4240642 · Received November 10, 2014

Report

Report Number
3009450871-2014-10595
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 22, 2014
Report Date
October 28, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK971544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE CONTROL UNIT WAS NOT FUNCTIONING. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE; A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Additional Manufacturer Narrative · 1

(B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE BATTERY REAMER DRILL DEVICE OVERHEATED AND STOPPED FUNCTIONING. ACCORDING TO THE REPORTER, THE DEVICE WAS THEN SWITCHED OFF AND ON AGAIN. HOWEVER, THE DEVICE WAS STILL NOT FUNCTIONING. THE REPORTER INDICATED THAT THE BATTERY INSIDE THE CASING WAS CHECKED AND IT WAS FOUND THAT PART OF THE BATTERY HOUSING WAS MELTED. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723368 BATT F/BPL 14.4VDC MOTOR, DRILL, PNEUMATIC HBB DEPUY SYNTHES POWER TOOLS D110325

Patients

Seq Age Sex Outcome Treatment
1