FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4240611 · Received November 10, 2014

Report

Report Number
2032227-2014-49876
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 11, 2014
Report Date
October 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. IT WAS ALSO REPORTED THAT THE INSULIN PUMP HAS AN UNRESPONSIVE KEYPAD. THE BLOOD GLUCOSE LEVELS WERE NOT INCLUDED IN THE REPORT. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724005 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 24 YR