FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240601 · Received November 10, 2014

Report

Report Number
2032227-2014-49867
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 15, 2014
Report Date
October 15, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO DISPLAY RAMPING ON ITS OWN ANOMALY NOTED. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS. UNIT RECEIVED WITH SEVERELY SCRATCHED LCD WINDOW.

Description of Event or Problem · 1

IT IS REPORTED THAT CUSTOMER EXPERIENCED KEYPAD ANOMALY. IT IS REPORTED THAT NUMBERS BEGIN RAMPING ON PUMP. CUSTOMER'S BLOOD GLUCOSE WAS 76 MG/DL. CUSTOMER WAS ADVISED THAT INSULIN PUMP WILL BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723320 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1