FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240598 · Received November 10, 2014

Report

Report Number
2032227-2014-31729
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 9, 2014
Report Date
October 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A NO DELIVERY ALARM AFTER A RESERVOIR CHANGE. CUSTOMER REPORTED THAT THE INSULIN PUMP MADE CLICKING AND GRINDING SOUNDS WHEN TRYING TO DELIVER INSULIN. CUSTOMER REPORTED HAVING BLOOD GLUCOSE LEVELS BETWEEN 100 TO 150 MG/DL DURING THE TIME OF THE ALARMS. DURING TROUBLESHOOTING, CUSTOMER WAS ABLE TO GET INSULIN TO EXIT WITHOUT A NO DELIVERY ALARM. BLOOD GLUCOSE LEVEL WAS 109 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723319 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAL

Patients

Seq Age Sex Outcome Treatment
1 72 YR