FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 4240567 · Received November 10, 2014

Report

Report Number
1416980-2014-39695
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K062457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS MANUFACTURED JULY 23, 2014 TO JULY 25, 2014. EVALUATION SUMMARY: THE SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION IDENTIFIED A SOLID WHITE PARTICLE APPROXIMATELY 0.86 MM IN LENGTH, FLOATING IN THE FLUID OF THE RESERVOIR. FOURIER TRANSFORM INFRARED SPECTROSCOPY DETERMINED THAT THE PARTICLE WAS POLYESTER. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. THIS ISSUE IS CURRENTLY BEING ADDRESSED THROUGH A CAPA. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HALF DAY INFUSOR WAS OBSERVED TO HAVE PARTICULATE MATTER INSIDE THE BALLOON DURING FILLING OF AN UNSPECIFIED SOLUTION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.