DEXTRUS 4136
Report
- Report Number
- 1028232-2014-004028
- Event Type
- Injury
- Date Received
- November 10, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 27, 2014
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVN
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
AS OF TODAY, THE MEDICAL DEVICE IS NOT AVAILABLE FOR ANALYSIS, THEREFORE THE DEVICE ITSELF COULD NOT BE INVESTIGATED. THE INFORMATION YOU PROVIDED HAS BEEN ENTERED INTO OUR QUALITY SYSTEM AS A COMPLAINT. THESE TYPES OF COMPLAINTS ARE USED TO EVALUATE SYSTEMS AND DEVICE PERFORMANCE THROUGHOUT OUR ORGANIZATION AND HELP TO MAINTAIN AND IMPROVE THE PERFORMANCE OF OUR DEVICES. SHOULD ADDITIONAL RELEVANT INFORMATION OR THE DEVICE ITSELF BECOME AVAILABLE, THE INVESTIGATION WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEXTRUS LEAD WAS EXHIBITING PACING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS, LOSS OF CAPTURE AND NOISE. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SUCCESSFULLY REPOSITIONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE LEAD REMAINS IN SERVICE AND NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723923 | DEXTRUS 4136 | PACER LEAD | NVN | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |