EON
Report
- Report Number
- 1627487-2014-26910
- Event Type
- Injury
- Date Received
- October 30, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 9, 2014
- Manufacturer
- ST. JUDE MED - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE #1 OF 2. REFERENCE MFR. REPORT:1627487-2014-26911. THE PATIENT REPORTS HE EXPERIENCED THE CHARGING SYSTEM BECOMING WARM DURING RECHARGING WHICH CAUSED A LITTLE DISCOMFORT IN HIS BACK. A REPLACEMENT MODEL 3726 CHARGING SYSTEM WAS SENT TO THE PATIENT. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 697090 | EON | SCS IPG | GZB | ST. JUDE MED - NEUROMODULATION | 3716 | 83499 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Other | IMPLANT DATE:| SCS LEAD: MODEL 3816 |