FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240458 · Received November 10, 2014

Report

Report Number
2032227-2014-49714
Event Type
Injury
Date Received
November 10, 2014
Date of Event
October 12, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 540 MG/DL AT THE TIME THE MOTOR ERROR OCCURRED AND 490 MG/DL WHEN THE ISSUE WAS BEING REPORTED. THE INSULIN PUMP HAD ALSO ALARMED NO DELIVERY. THE CUSTOMER WAS NOT USING THE SENSOR FEATURE ON THE INSULIN PUMP. THE REWIND SEQUENCE WAS COMPLETED AND IT WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722696 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 10 YR