PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-49713
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- July 1, 2014
- Report Date
- October 12, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP HAD BROKEN BATTERY TUBE THREADS, A CRACKED DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A BROKEN RESERVOIR TUBE LIP, SCRATCHED RESERVOIR TUBE WINDOW, A CRACKED BELT CLIP SLOT, A MISSING END CAP STICKER AND A STAINED ADDRESS AND SERIAL NUMBER LABEL.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4)
IT WAS REPORTED THAT THE BATTERY CAP FOR THE CUSTOMER'S INSULIN PUMP WAS STRIPPED AND PARTS OF IT WERE MISSING. THE CUSTOMER STATED SHE COULD NOT GET THE INSULIN PUMP TO TURN ON DUE TO THE DAMAGED BATTERY CAP. HER BLOOD GLUCOSE WAS 263 MG/DL. THE CUSTOMER STATED THE BATTERY CAP WOULD NOT SCREW ON AND WOULD NOT POWER ON AS A RESULT. THE CUSTOMER STATED THAT THREADING ON THE BATTERY CAP WAS MISSING AND THERE WERE CRACKS AND PARTS MISSING IN THE BATTERY COMPARTMENT. THE CUSTOMER STATED SHE WAS KEEPING THE INSULIN PUMP IN HER BRA AND THE DAMAGE COULD BE DUE TO PERSPIRATION. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722221 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |