FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240456 · Received November 10, 2014

Report

Report Number
2032227-2014-49712
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 12, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A BROKEN SOLDER JOINT ON THE INTERFACE BOARD. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP STOPPED WORKING. SHE STATED SHE REPLACED THE BATTERY AND NOTHING CAME BACK UP. HER BLOOD GLUCOSE WAS 130 MG/DL. THE CUSTOMER STATED THE INSULIN PUMP HAD NOT BEEN BUMPED, DROPPED, OR EXPOSED TO MOISTURE AND THERE WAS NO APPARENT DAMAGE. THE CUSTOMER INSERTED A NEW BATTERY AND THE DISPLAY DID NOT RETURN. AFTER THE CUSTOMER WAS ADVISED TO REMOVE THE BATTERY FOR 10 MINUTES AND CLEAN THE BATTERY CAPT CONTACTS, ALLOW TO DRY, AND INSERT A NEW BATTERY, THE CUSTOMER STATED THE DISPLAY DID NOT RETURN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721752 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 44 YR