PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-49712
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 12, 2014
- Report Date
- October 12, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A BROKEN SOLDER JOINT ON THE INTERFACE BOARD. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.
THE CUSTOMER REPORTED THAT THE INSULIN PUMP STOPPED WORKING. SHE STATED SHE REPLACED THE BATTERY AND NOTHING CAME BACK UP. HER BLOOD GLUCOSE WAS 130 MG/DL. THE CUSTOMER STATED THE INSULIN PUMP HAD NOT BEEN BUMPED, DROPPED, OR EXPOSED TO MOISTURE AND THERE WAS NO APPARENT DAMAGE. THE CUSTOMER INSERTED A NEW BATTERY AND THE DISPLAY DID NOT RETURN. AFTER THE CUSTOMER WAS ADVISED TO REMOVE THE BATTERY FOR 10 MINUTES AND CLEAN THE BATTERY CAPT CONTACTS, ALLOW TO DRY, AND INSERT A NEW BATTERY, THE CUSTOMER STATED THE DISPLAY DID NOT RETURN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 721752 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |