FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4240392 · Received November 10, 2014

Report

Report Number
2032227-2014-49662
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 11, 2014
Report Date
October 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH A CRACKED DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNER, A CRACKED RESERVOIR TUBE LIP AND A SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR AND WAS GIVING BATTERY ISSUES. CUSTOMER STATED HE RECEIVED A NEW BATTERY CAP AND ISSUES PERSISTED. CUSTOMER'S BLOOD GLUCOSE WAS 110 MG/DL. CUSTOMER DOES NOT RECALL ANY PREVIOUS SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE DEVICE TO ALARM. THE DEVICE WAS NOT EXPOSED TO MRI. ALARM WAS CLEARED. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE. CUSTOMER DOES USE THE SENSOR FEATURE. CUSTOMER WAS ADVISED ABOUT FALSE MOTOR ERROR ALARMS OCCURRING, IF CUSTOMER TRIED TO SEE THE SENSOR GLUCOSE GRAPH WHILE BOLUSING. CUSTOMER WAS ABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722504 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 25 YR