PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-49661
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 11, 2014
- Report Date
- October 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
NO UNEXPECTED RESETS NOTED DURING TESTING. THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND AND BASIC OCCLUSION TESTS. THE MOTOR ERROR ALARMED DURING PRIME TEST DUE TO MOISTURE DAMAGE ON MOTOR ASSEMBLY. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS AND MISSING END CAP STICKER.
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THE INSULIN PUMP JUST STOPPED WORKING AND IT ALARMED MOTOR ERROR. CUSTOMER DID SELF TEST AND DEVICE STILL ALARMED. CUSTOMER'S BLOOD GLUCOSE WAS 220 MG/DL. ALARM OCCURRED DURING BOLUS DELIVERY THE FIRST TIME AND THEN IT OCCURRED WHEN FILLING TUBING. THE DEVICE RESET IT SELF AND CUSTOMER HAD RESET BASAL RATES AND TIME. THE DEVICE WAS NOT EXPOSED TO MRI. ALARM WAS CLEARED. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE. CUSTOMER DOES NOT USE THE SENSOR FEATURE. CUSTOMER WAS UNABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722452 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |