FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4240360 · Received November 10, 2014

Report

Report Number
2032227-2014-49489
Event Type
Injury
Date Received
November 10, 2014
Date of Event
November 3, 2014
Report Date
November 3, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT FUNCTIONED PROPERLY. NO DELIVERY ANOMALY WAS NOTED DURING TESTING. ALL BOLUSES WERE DELIVERED AND PROPERLY RECORDED IN BOLUS HISTORY. UNIT RECEIVED WITH MINOR SCRATCHED LCD WINDOW AND CRACKED RESERVOIR TUBE LIP. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT HIGH BLOOD GLUCOSE LEVELS. CUSTOMER'S BLOOD GLUCOSE READING WAS 429 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721960 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 42 YR