PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-49836
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 11, 2014
- Report Date
- October 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP BUTTONS ALL RESPONDED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS FOUND AT THE KEYPAD TRACES. NO MOISTURE DAMAGE WAS FOUND AT THE ELECTRONIC OR MOTOR ASSEMBLIES DURING THE VISUAL INSPECTION. NO BUTTON ERROR ALARMS WERE NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED RESERVOIR TUBE LIP, A CRACKED BELT CLIP SLOT, AND MINOR SCRATCHES ON THE DISPLAY WINDOW.
CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A BUTTON ERROR ALARM. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD RECENTLY BEEN EXPOSED TO RAIN. BLOOD GLUCOSE LEVEL WAS 179 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 721729 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |