FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4240139
·
Received November 10, 2014
Report
- Report Number
- 1031452-2014-18188
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Report Date
- October 17, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
THE INITIAL OXYGEN READING WAS 68-69% WHICH SHOULD HAVE PRODUCED AN "ALARM" BUT WASN'T MENTIONED WITH INITIAL COMPLAINT. THE REPAIR STATEMENT INDICATES THAT A LOOSE HEAT EXCHANGER TUBING CLAMP, SIEVE BEDS TUBING AND PE VALVE TUBING CLAMP WERE LEADING TO THE LOW O2 READING.