PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4240093 · Received November 10, 2014

Report

Report Number
2032227-2014-49813
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 11, 2014
Report Date
October 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.(B)(4)

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED AND LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE NEAR THE RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS, AND A BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM AND HAD LOOSE DRIVE SUPPORT CAP. TROUBLESHOOTING WAS DONE. CUSTOMER'S BLOOD GLUCOSE WAS 143 MG/DL. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.