FDA Adverse Event Malfunction Summary report: N

MAQUET HCU20

MDR report key: 4240038 · Received November 3, 2014

Report

Report Number
8010762-2014-01098
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 4, 2014
Report Date
October 4, 2014
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DWC
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A MAQUET SERVICE TECH INVESTIGATED THE DEVICE AND FOUND THE FAILURE WITH TANK PT (B)(4). THE PT (B)(4) WAS REPLACED AND THE UNIT WAS TESTED TO FACTORY SPECS. DEVICE WAS RETURNED TO SERVICE. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED IF ADD'L INFO BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT START-UP, THE DEVICE DISPLAYED TANK TEMPERATURE UNK VALUES. DISPLAY SHOWED COMP ERROR. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703495 MAQUET HCU20 CONTROLLER, TEMP, CARDIOPULMONARY BYPASS DWC MAQUET CARDIOPULMONARY AG

Patients

Seq Age Sex Outcome Treatment