FDA Adverse Event Malfunction Summary report: N

NOMAD PRO

MDR report key: 4240007 · Received November 3, 2014

Report

Report Number
3005417494-2014-00003
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
September 8, 2014
Report Date
October 7, 2014
Manufacturer
ARIBEX, INC.
Product Code
EHD
PMA / PMN Number
K081664
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE HANDSET WAS RECEIVED AND INSPECTED. OUR RECORDS INDICATE THAT THE HANDSET AND THE DEVICE WERE ASSEMBLED (B)(6) 2012. EXPECTED END OF LIFE FOR THE BATTERY IS 300 CHARGE/DISCHARGE CYCLES AS PER THE USER MANUAL. THE UNIT WAS INSPECTED AND IT WAS DETERMINED THAT THE FAILURE WAS DUE TO AN OVER-DISCHARGE OF THE BATTERY THAT LED TO A THERMAL EVENT. THE HEAT FROM THE BATTERY FAILURE LED TO SMALL BURN MARKS ON THE INSIDE BOTTOM OF THE HANDSET. THE BATTERY HANDSET ENCLOSURE WAS NOT COMPROMISED. THERE WERE NO INJURIES REPORTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE BATTERY PACK (HANDSET) OF THE UNIT STARTED SMOKING ON CHARGING CRADLE. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
704039 NOMAD PRO EXTRAORAL SOURCE X-RAY SYSTEM EHD ARIBEX, INC. FF-0005

Patients

Seq Age Sex Outcome Treatment
1