FDA Adverse Event
Malfunction
Summary report: N
NOMAD PRO
MDR report key: 4240007
·
Received November 3, 2014
Report
- Report Number
- 3005417494-2014-00003
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- September 8, 2014
- Report Date
- October 7, 2014
- Manufacturer
- ARIBEX, INC.
- Product Code
- EHD
- PMA / PMN Number
- K081664
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE HANDSET WAS RECEIVED AND INSPECTED. OUR RECORDS INDICATE THAT THE HANDSET AND THE DEVICE WERE ASSEMBLED (B)(6) 2012. EXPECTED END OF LIFE FOR THE BATTERY IS 300 CHARGE/DISCHARGE CYCLES AS PER THE USER MANUAL. THE UNIT WAS INSPECTED AND IT WAS DETERMINED THAT THE FAILURE WAS DUE TO AN OVER-DISCHARGE OF THE BATTERY THAT LED TO A THERMAL EVENT. THE HEAT FROM THE BATTERY FAILURE LED TO SMALL BURN MARKS ON THE INSIDE BOTTOM OF THE HANDSET. THE BATTERY HANDSET ENCLOSURE WAS NOT COMPROMISED. THERE WERE NO INJURIES REPORTED.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE BATTERY PACK (HANDSET) OF THE UNIT STARTED SMOKING ON CHARGING CRADLE. NO INJURIES WERE REPORTED.