REHAB CONT AIR-LIGHT R
Report
- Report Number
- 2085446-2012-00007
- Event Type
- Malfunction
- Date Received
- October 30, 2014
- Date of Event
- September 4, 2012
- Report Date
- October 24, 2014
- Manufacturer
- L&L INTERNATIONAL CO LTD
- Product Code
- ITQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE PRODUCT HIGNE IS SET IN -15 DEGREES AND 105 DEGREE FLEXION. THIS INDICATES THE FULL RANGE OF MOTION OF THE BRACE HAS BEEN THE SETTING WHEN FAILED. FURTHERMORE, THE FLEXION LIMITER HAVE THE PLASTIC PLUGS IN THEM WHILE THE EXTENSION LIMITERS DON'T. THE PRODUCT IS VERY HEAVILY WORN, PADDING AND STRAPS ARE WORN A LOT. THE TWO THIGH STRAPS ARE TORN THROUGH THE ULTRASONIC WELD OF THE HOOK PART TO THE STRAP. TOP STRAP HAS DELAMINATED THROUGH HIGH LOADING AS THE FIBRES IN THE STRAP INDICATED. THE LOWER THIGH STRAP HAS TORN AND DELAMINATED, PROBABLY AT LOWER THIGH STRAP HAS TORN AND DELAMINATED, PROBABLY AT THE SAME TIME. THERE STRAPS SHOULD NOT GIVE IN UNDER THESE LOADINGS AND ESPECIALLY AT THE SETTINGS GIVEN IN THE BRACE. THE FLEXION AND EXTENSION LIMITERS ARE INTACT AND SHOW NO SIGN OF FAILURE. NO SIGN OF HIGH LOADING CAN BEN SEEN ON THE HINGE CENTER. THE MARKS THERE ARE MORE AS A SIGN OF NORMAL WEAR AND TEAR. IT WOULD SEEM VERY UNLIKELY THAT THE UNIT WOULD FAIL DRASTICALLY UNLESS THE USER TRIPPED OR OTHERWISE WAS IN ABNORMAL CONDITIONS BASED ON THE ASSESSMENT OF THE BRACE.
THE PT WAS WEARING THE REHAB KNEE BRACE WHILE WALKING UP A FLIGHT OF STAIRS. THE PT CLAIMS THE VELCRO ON THE BRACE FAILED CAUSING HIM TO FALL. HE WAS ABLE TO CATCH AND BRACE HIMSELF BUT HE LANDED ON HIS KNEE AND FRACTURED HIS PATELLA. HE WENT IN FOR SURGERY THE NEXT DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 695121 | REHAB CONT AIR-LIGHT R | JOINT, KNEE, EXTERNAL BRACE | ITQ | L&L INTERNATIONAL CO LTD | B-220500000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |