FDA Adverse Event Summary report: N

3M ESPE PENTAMIX 3 AUTOMATIC MIXING UNIT

MDR report key: 4236985 · Received October 31, 2014

Report

Report Number
9611385-2014-00014
Date Received
October 31, 2014
Report Date
September 11, 2014
Manufacturer
3M DEUTSCHLAND GMBH
Product Code
EFD
PMA / PMN Number
K991913
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR AND ANALYZED. IT WAS FOUND, THAT THE DEVICE HAD A BROKEN HOUSING. INSIDE THE MACHINE A MELTED PART WAS FOUND. POTENTIALLY A SHORT CIRCUIT CAUSED OVERHEATING AND MELTING. THE CAUSE OF THE BROKEN HOUSING COULDN'T BE DETERMINED, BUT MOST LIKELY IMPROPER HANDLING OR BEING DROPPED LED TO THE DAMAGE. HOWEVER, THE ANALYSIS SHOWED THAT IT IS UNLIKELY THAT THE BROKEN HOUSING IS RELATED TO THE EVENT REPORTED. THE PENTAMIX DEVICE IS DESIGNED IN A WAY THAT NO ELECTRICALLY CONDUCTING PARTS ARE EXPOSED TO THE DEVICE SURFACE. THE HOUSING IS MADE COMPLETELY OF PLASTICS. THE ONLY PARTS MADE OF METAL EXPOSED TO THE OUTSIDE ARE THE CONNECTORS AND ONE SMALL PART OF THE DRIVESHAFT. THESE PARTS ARE GROUNDED. SO DURING INTENDED USE THERE'S NO POSSIBILITY TO GET A LIFE-THREATENING ELECTRICAL SHOCK BY TOUCHING THE OUTSIDE OF THE DEVICE. AS THE USER TOUCHED OUTSIDE PARTS OF THE DEVICE AND NO CONDUCTING PARTS, IT IS NOT CLEAR WHERE THE ELECTRICAL SHOCK CAME FROM. IT CAN ALSO BE RELATED TO A STATIC CHARGE OF THE USER.

Description of Event or Problem · 1

THE 3M ESPE WAS INFORMED ON (B)(6) 2014 THAT A DENTAL OFFICE EMPLOYEE RECEIVED AN ELECTRICAL SHOCK AFTER UNPLUGGING A 3M ESPE PENTAMIX MACHINE FROM A POWER OUTLET. IT WAS REPORTED THAT THE EMPLOYEE WAS UNPLUGGING THE MACHINE TO SEND IT TO THE REPAIR SHOP WHEN SHE RECEIVED THE SHOCK; NO INJURY WAS SUSTAINED, NO MEDICAL ATTENTION WAS SOUGHT, AND THE EMPLOYEE IS CURRENTLY FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
698231 3M ESPE PENTAMIX 3 AUTOMATIC MIXING UNIT AMALGAMATOR, DENTAL, AC-POWERED EFD 3M DEUTSCHLAND GMBH

Patients

Seq Age Sex Outcome Treatment
1 Other