FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4234118 · Received November 7, 2014

Report

Report Number
1031452-2014-17836
Event Type
Malfunction
Date Received
November 7, 2014
Report Date
October 15, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER INDEPENDENT REPAIR CENTER STATEMENT: ALLEGED ALARMING THE KEY FAILURE IS THE SIEVE BEDS NOT WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
718002 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other