FDA Adverse Event
Malfunction
Summary report: N
OT VERIO IQ METER
MDR report key: 4233990
·
Received November 7, 2014
Report
- Report Number
- 2939301-2014-29876
- Event Type
- Malfunction
- Date Received
- November 7, 2014
- Report Date
- October 29, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 ¿ (12/30/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 12/3/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 12/18/2014 WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A DEFECTIVE CAPACITOR C85. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
Description of Event or Problem · 1
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 1. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 718319 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3667525 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |