FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4233840 · Received November 6, 2014

Report

Report Number
2531779-2014-31777
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 22, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/22/2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD COVER WAS FOUND TO BE PEELING UP AT THE OK BUTTON. DURING TESTING, THE OK KEYPAD BUTTON WAS FOUND TO BE INTERMITTENTLY UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE OK KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE. INVESTIGATION ALSO REVEALED THAT THERE WAS CONTAMINATION UNDER ALL KEY CONTACTS. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/22/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715909 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 25 YR