PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 4233777 · Received November 6, 2014

Report

Report Number
2024168-2014-07287
Event Type
Injury
Date Received
November 6, 2014
Date of Event
October 13, 2014
Report Date
October 14, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT WAS REPORTED THAT THE PROGLIDE DEVICE WAS USED IN A CALCIFIED VESSEL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. THE DEVICE WAS REPORTED TO BE DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER PERCLOSE PROGLIDE DEVICE REFERENCED IS FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT ARTERIOTOMY CLOSURE OF THE HEAVILY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH PROGLIDE DEVICES, USING A PRE-CLOSE TECHNIQUE, VIA A 6F SHEATH PRIOR TO AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, CUFF MISSES OCCURRED WITH BOTH DEVICES. TWO ADDITIONAL PROGLIDE DEVICES WERE USED FOR SUCCESSFUL SUTURE PLACEMENT. THE SHEATH WAS UPSIZED TO 13F AND THE AAA PROCEDURE WAS COMPLETED. THE PREPLACED SUTURES OF THE THIRD AND FOURTH PROGLIDE DEVICES WERE USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. IT WAS REPORTED THAT THE PHYSICIAN IS TRAINED IN THE USE OF THE PROGLIDE DEVICE AND TRAINED IN THE LARGE HOLE TECHNIQUE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714685 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 40709K1

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention SHEATH: 6F, 13F