FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233663 · Received November 6, 2014

Report

Report Number
2032227-2014-48928
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 9, 2014
Report Date
June 24, 2016
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY; MOISTURE DAMAGE WAS ALSO FOUND ON THE MOTOR ASSEMBLY. UNABLE TO PERFORM FUNCTIONAL CHECK INCLUDING REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, THE EXCESSIVE NO DELIVERY, DISPLACEMENT, SELF-TEST, A21 TEST AND ALL OPERATING CURRENT TEST DUE TO BLANK DISPLAY. NO UNEXPECTED AUDIO BEEP NOTED DURING OUR TESTING. THE INSULIN PUMP HAD MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND SCRATCHED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BEEP ANOMALY AND A BLANK DISPLAY. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS NOT REPORTED. THE CUSTOMER WAS ADVISED TO REVERT TO BACK UP PLAN. THE PRODUCT WILL BE RETURNED. NOTHING FURTHER TO REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715507 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754LWWS

Patients

Seq Age Sex Outcome Treatment
1