FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233622 · Received November 6, 2014

Report

Report Number
2032227-2014-48782
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
March 20, 2014
Report Date
October 10, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTER. UNABLE TO VERIFY NO DELIVERY ALARM DUE TO PRIME ANOMALY. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULIN PUMP WAS GIVING AN UNUSUAL NO DELIVERY ERROR ALARM. CUSTOMER STATED THAT AFTER REWINDING THE CUSTOMER INSERTED THE RESERVOIR AND PRESSED ACT. THEN THE DEVICE STARTED COUNTING INSULIN UNITS AND THEN GAVE A NO DELIVERY ALARM. STATED THE DEVICE WAS NOT WORKING. CUSTOMER'S BLOOD GLUCOSE WAS NOT PROVIDED. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715931 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWS

Patients

Seq Age Sex Outcome Treatment
1