PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-48742
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- September 25, 2014
- Report Date
- October 9, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(2).
NO MOTOR ERROR ALARM NOTED. THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO MOISTURE DAMAGE ON FORCE SENSOR. THE INSULIN PUMP WAS UNABLE TO PERFORM EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST DUE TO PRIME/FILL ANOMALY. THE MOTOR WAS TESTED OUTSIDE THE INSULIN PUMP AND PASSED. THE INSULIN PUMP HAD A CRACKED RESERVOIR TUBE, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BELT CLIP SLOT, BROKEN BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON LCD WINDOW.
CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THERE IS A MOTOR ERROR ALARM. THE BLOOD GLUCOSE READING IS BETWEEN 160-200 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 714697 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |