FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233549 · Received November 6, 2014

Report

Report Number
2032227-2014-48742
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
September 25, 2014
Report Date
October 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(2).

Additional Manufacturer Narrative · 1

NO MOTOR ERROR ALARM NOTED. THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO MOISTURE DAMAGE ON FORCE SENSOR. THE INSULIN PUMP WAS UNABLE TO PERFORM EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST DUE TO PRIME/FILL ANOMALY. THE MOTOR WAS TESTED OUTSIDE THE INSULIN PUMP AND PASSED. THE INSULIN PUMP HAD A CRACKED RESERVOIR TUBE, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BELT CLIP SLOT, BROKEN BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON LCD WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THERE IS A MOTOR ERROR ALARM. THE BLOOD GLUCOSE READING IS BETWEEN 160-200 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714697 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAP

Patients

Seq Age Sex Outcome Treatment
1 15 YR