FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233530 · Received November 6, 2014

Report

Report Number
2032227-2014-48795
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
March 10, 2014
Report Date
March 10, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. NO MOISTURE DAMAGE ON ELECTRONICS NOTED. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

CUSTOMER'S NURSE REPORTED THAT THE INSULIN PUMP RETURNED A BUTTON ERROR ALARM. CUSTOMER REPORTED THAT THE INSULIN PUMP MAY HAVE RECENTLY BEEN EXPOSED TO SWEAT. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.